DocumentCode :
1320065
Title :
Experimental Data for Locating VSM Pickup Coils and Minimizing the Image Effect
Author :
Lorenz, B.E. ; Graham, C.D.
Author_Institution :
Dept. of Electr. Eng., Widener Univ., Chester, PA, USA
Volume :
48
Issue :
11
fYear :
2012
Firstpage :
2914
Lastpage :
2916
Abstract :
From measurements of the magnetic field produced in the air gap of an electromagnet by a small permanent magnet sample, we can predict the relative sensitivity of a vibrating-sample magnetometer (VSM) for any practical location of the pickup coils. By repeating the measurements with the pole pieces removed, we can simulate the case of complete magnetic saturation (permeability μ = 1) of the pole pieces, and thus predict the relative magnitude of the image effect for any coil location. Under some conditions, a coil position can be chosen that eliminates the image effect.
Keywords :
coils; electromagnets; magnetic field measurement; magnetic sensors; magnetometers; permanent magnets; pick-ups; vibration measurement; VSM pickup coil location; air gap; electromagnet; image effect minimization; magnetic field measurement; magnetic saturation; permanent magnet sample; pole piece removal; vibrating-sample magnetometer; Coils; Electromagnets; Magnetic resonance imaging; Magnetometers; Permanent magnets; Permeability; Saturation magnetization; Image effect; magnetic measurements; vibrating-sample magnetometer (VSM);
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/TMAG.2012.2198620
Filename :
6332571
Link To Document :
بازگشت