DocumentCode
1322393
Title
Examination of Precise Measurement of DC Magnetic Properties of Permalloy Under Low Flux Density More Than a Few mT
Author
Mimura, M. ; Takahashi, N. ; Nakano, M. ; Ujigawa, S. ; Shinnoh, T. ; Miyagi, D.
Author_Institution
Dept. Electr. & Electron. Eng., Okayama Univ., Okayama, Japan
Volume
48
Issue
11
fYear
2012
Firstpage
3614
Lastpage
3617
Abstract
Permalloy is used in a magnetically shielded room to inhibit the penetration of external magnetic field, etc. The flux density in permalloy used in the shielded room is of the order of a few mT or lower. In order to examine the performance of such a shielded room, the dc magnetic properties at low flux density is necessary. In this paper, we developed a measurement method of dc (5 mHz) magnetic properties of permalloy under such a low flux density. It is shown that the dc magnetic properties could be measured more than 4 mT by paying the attention to the degaussing of specimen and applying the correction of drift and offset of measurement system.
Keywords
Permalloy; magnetic field effects; magnetic field measurement; magnetic flux; magnetic shielding; DC magnetic properties; external magnetic field; low flux density; magnetically shielded room; permalloy; precise measurement; specimen degaussing; Current measurement; Density measurement; Frequency measurement; Magnetic flux; Magnetic hysteresis; Magnetic noise; Magnetic shielding; DC magnetic property; low flux density; permalloy;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/TMAG.2012.2199471
Filename
6332951