Title :
High-frame-rate echocardiography with reduced sidelobe level
Author :
Hasegawa, Hiroshi ; Kanai, Hiroshi
Author_Institution :
Grad. Sch. of Biomed. Eng., Tohoku Univ., Sendai, Japan
Volume :
59
Issue :
11
fYear :
2012
fDate :
11/1/2012 12:00:00 AM
Abstract :
Echocardiography has become an indispensable modality for diagnosis of the heart. It enables observation of the shape of the heart and estimation of global heart function based on B-mode and M-mode imaging. Methods for echocardiographic estimation of myocardial strain and strain rate have also been developed to evaluate regional heart function. Furthermore, it has been recently shown that echocardiographic measurements of transmural transition of myocardial contraction/ relaxation and propagation of vibration caused by closure of the heart valve would be useful for evaluation of myocardial function and viscoelasticity. However, such measurements require a frame rate (typically >;200 Hz) much higher than that achieved by conventional ultrasonic diagnostic equipment. We have recently realized a high frame rate of about 300 Hz with a full field of view of 90° using diverging transmit beams and parallel receive beamforming. Although high-frame-rate imaging was made possible by this method, the side lobe level was slightly larger than that of the conventional method. To reduce the side lobe level, phase coherence imaging has recently been developed. Using this method, the spatial resolution is improved and the side lobe level is also reduced. However, speckle-like echoes, for example, echoes from the inside of the heart wall, are also suppressed. In the present study, a method for reducing the side lobe level while preserving speckle-like echoes was developed. The side lobe level was evaluated using a wire phantom. The side lobe level of the high-frame-rate imaging using unfocused diverging beams was improved by 13.3 dB by the proposed method. In in vivo measurements, a B-mode image of the heart of a 23-year-old healthy male could be obtained while preserving the speckle pattern in the heart wall at a frame rate of 316 Hz with a full field of view of 90°.
Keywords :
biomechanics; echocardiography; estimation theory; image resolution; medical image processing; muscle; phantoms; speckle; ultrasonic imaging; ultrasonic propagation; viscoelasticity; B-mode imaging; M-mode imaging; age 23 yr; conventional ultrasonic diagnostic equipment; echocardiographic estimation; echocardiographic measurements; frequency 316 Hz; global heart function estimation; heart diagnosis; heart valve; high-frame-rate echocardiography; image contrast; image resolution; myocardial contraction-relaxation; myocardial strain rate; parallel receive beamforming; phase coherence imaging; reduced sidelobe level; regional heart function evaluation; spatial resolution; speckle pattern; speckle-like echoes; transmural transition; unfocused diverging beams; vibration propagation; viscoelasticity; wire phantom; Acoustics; Array signal processing; Coherence; Delay effects; Heart; Imaging; Silicon; Echocardiography; Humans; Image Processing, Computer-Assisted; Male; Phantoms, Imaging; Young Adult;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2012.2490
Filename :
6343284
Link To Document :
بازگشت