Title : 
Suppression of Artifacts in Biomeasurement System by Pneumatic Filtering
         
        
            Author : 
Kurihara, Yosuke ; Watanabe, Kajiro
         
        
            Author_Institution : 
Dept. of Comput. & Inf. Sci., Seikei Univ., Tokyo, Japan
         
        
        
        
        
            fDate : 
3/1/2012 12:00:00 AM
         
        
        
        
            Abstract : 
The pneumatic biomeasurement system is a noninvasive, constraint-free method for monitoring the heartbeat, respiration and other vital signs of a person in bed. Body microvibrations transmitted through the bed cushion act on the air in a thin air mattress under the cushion. Changes in air pressure synchronized with the microvibrations are detected by a pressure sensor. However, the high sensitivity of the sensor results in output saturated with body motion artifacts. This paper describes a novel method for pneumatic filtering of these artifacts. A theoretical model was developed as the basis for designing a pneumatic filter to reduce body motion artifacts. The validity of the proposed high-pass filtering method was examined through experiments.
         
        
            Keywords : 
cardiology; high-pass filters; patient monitoring; physiological models; pneumatic systems; pneumodynamics; pressure sensors; artifact suppression; bed cushion act; body microvibrations; body motion artifacts; constraint-free method; heartbeat monitoring; high-pass filtering method; pneumatic biomeasurement system; pneumatic filter; pneumatic filtering; pressure sensor; respiration; thin air mattress; Band pass filters; Electron tubes; Frequency measurement; Heart beat; Monitoring; Pressure measurement; Body motion; constraint-free biosensing; heartbeat; respiration;
         
        
        