  • Abstract
    The development of 2-D array transducers has received much recent interest. Unfortunately, fabrication of high density 2-D arrays is difficult due to the large number of electrical interconnections which must be made to the back side of the elements. A typical array operating at 2.2 MHz may have 256 or more connections within a 16.4 mm circular footprint. Interconnection becomes even more challenging as operating frequencies increase. To solve this problem, we have developed a multilayer flexible (MLF) circuit interconnect consisting of a polyimide dielectric with inter-laminar vias routing signals vertically and etched metal traces routing signals horizontally. A transducer is fabricated from an MLF by bonding a PZT chip to its surface and dicing the chip into individual elements, with the saw kerf extending partially into the top polyimide layer to ensure physical and electrical isolation of the elements. The KLM model was used to compare the performance of an MLF 2-D array to a conventional hand wired 2-D array. MLF and wire guide transducers were fabricated, each with 256 active elements, 0.4 mm interelement spacing, and 2.2 MHz center frequency. Vector impedance, pulse length, bandwidth, angular response, and cross-coupling were found to be comparable in both types of arrays. Using the MLF, however, fabrication time was reduced dramatically. More importantly, MLF technology may be used to increase 2-D array connection density beyond the limitations of current of hand wired fabrication techniques. Thus MLF circuits provide a means for the interconnection of current and future high frequency 2-D arrays.
  • Keywords
    biomedical ultrasonics; ultrasonic transducer arrays; wiring; 2.2 MHz; PZT chip; angular response; array connection density; array transducers; bandwidth; cross-coupling; electrical isolation; etched metal traces; fabrication time; inter-laminar vias; interelement spacing; medical ultrasound; multilayer flexible circuit interconnection; polyimide dielectric; pulse length; two-dimensional arrays; vector impedance; wire guide transducers; Biomedical transducers; Fabrication; Frequency; Integrated circuit interconnections; Nonhomogeneous media; Polyimides; Routing; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.660144
  • Filename
    660144