• DocumentCode
    1337484
  • Title

    High-power calibration and measurement method for bio-electromagnetic study

  • Author

    Liu, Xindong ; Chen, H.-J. ; Chen, Xia ; Wen, Di ; Parini, Clive

  • Author_Institution
    Beijing University of Posts and Telecommunications,People??s Republic of China
  • Volume
    6
  • Issue
    6
  • fYear
    2012
  • fDate
    11/1/2012 12:00:00 AM
  • Firstpage
    420
  • Lastpage
    426
  • Abstract
    A novel method for high-power calibration/measurement is presented in this work. This method employs a coupler and a dipole antenna to achieve high-power measurement. The coupler serves as a low-power sensor, whereas the dipole antenna is responsible for the high-power measurement. This method successfully calibrated a 600 W 13.56 MHz radio frequency generator and is potentially useful for even higher power measurement, as well as for hazardous high-intensity electromagnetic environment measurement, avoiding the use of expensive high-power metres.
  • fLanguage
    English
  • Journal_Title
    Science, Measurement & Technology, IET
  • Publisher
    iet
  • ISSN
    1751-8822
  • Type

    jour

  • DOI
    10.1049/iet-smt.2011.0187
  • Filename
    6356015