• DocumentCode
    1344928
  • Title

    Electrode design for coplanar-grid detectors

  • Author

    Luke, P.N. ; Amman, M. ; Prettyman, T.H. ; Russo, P.A. ; Close, D.A.

  • Author_Institution
    Ernest Orlando Lawrence Berkeley Nat. Lab., CA, USA
  • Volume
    44
  • Issue
    3
  • fYear
    1997
  • fDate
    6/1/1997 12:00:00 AM
  • Firstpage
    713
  • Lastpage
    720
  • Abstract
    The coplanar-grid charge sensing technique provides a method for improving the spectral response of gamma-ray detectors based on compound semiconductors, which typically have poor charge transport properties. The technique functions by effectively modifying the charge induction characteristics of the detector such that the dependence of detector signal on the depth of radiation interaction is minimized. The effectiveness of this technique however can be compromised by non-uniform charge induction characteristics across the detector. This paper examines such non-uniformity due to fringe effects near the detector edges. Alternate electrode configurations are studied that provide effective compensation for such effects. Results from experimental measurements and computer simulations are presented
  • Keywords
    gamma-ray detection; semiconductor counters; charge induction characteristics; charge transport properties; compound semiconductors; coplanar-grid charge sensing technique; coplanar-grid detectors; electrode design; fringe effects; gamma-ray detectors; spectral response; Computer simulation; Electrodes; Energy resolution; Gamma ray detectors; Laboratories; Radiation detectors; Semiconductor radiation detectors; Statistics; Temperature; Voltage;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.603739
  • Filename
    603739