Title :
Effect of electrode size on impedance images of two- and three-dimensional objects
Author :
Newell, J.C. ; Peng, Y. ; Edic, P.M. ; Blue, R.S. ; Jaia, H. ; Newell, R.T.
Author_Institution :
Dept. of Biomed. Eng., Rensselaer Polytech. Inst., Troy, NY, USA
fDate :
4/1/1998 12:00:00 AM
Abstract :
The sensitivity of an impedance imaging system to small cylindrical inhomogeneities in two-dimensional (2-D) and three-dimensional (3-D) saline tanks was studied for different height electrodes and different height targets. Experimental results were compared with analytical models. Inhomogeneities in the 3-D tank having limited vertical extent were detected by electrodes of vertical size comparable to that of the inhomogeneity. Taller electrodes had increased sensitivity to short targets to only a limited extent. When the electrode height was more than twice that of the target, sensitivity decreased or remained the same with further increases in electrode height. The system was less sensitive to inhomogeneities in the 3-D tank than to those in the 2-D tank. The distinguishability of conductors was greater than that of insulators in the 3-D tank, and the opposite was true in the 2-D tank, consistent with an analytical result.
Keywords :
electric impedance imaging; electrodes; 2-D tank; 3-D tank; analytical models; conductors distinguishability; electrode size effect; impedance images; impedance imaging system sensitivity; medical diagnostic imaging; saline tanks; small cylindrical inhomogeneities; three-dimensional objects; two-dimensional objects; Analytical models; Biomedical electrodes; Biomedical engineering; Biomedical imaging; Biomedical measurements; Conductivity; Conductors; Copper; Impedance; Two dimensional displays; Agar; Copper; Electric Impedance; Electrodes; Image Processing, Computer-Assisted; Materials Testing; Polyvinyl Chloride; Sensitivity and Specificity; Tomography;
Journal_Title :
Biomedical Engineering, IEEE Transactions on