• DocumentCode
    1352027
  • Title

    Activity based neural front-end recording system

  • Author

    Aslam, J. ; Merken, P. ; Huys, Raoul ; Akif Erismis, M. ; Firat Yazicioglu, Refet ; Puers, Robert ; Van Hoof, Chris

  • Author_Institution
    IMEC, Leuven, Belgium
  • Volume
    47
  • Issue
    21
  • fYear
    2011
  • Firstpage
    1170
  • Lastpage
    1172
  • Abstract
    Presented is a new method for implementing sensor arrays for in vitro recording of neural action potentials. A key feature of the system is that it is divided into two parts, i.e. neural activity detection and neural recording, which softens key design trade-offs such as noise, power, sensor pitch, and observability. The term observability is defined as the percentage of the total sensors that can be monitored in real time. This system should provide high observability, low noise, low power and reasonable pitch with all these key specifications within acceptable limits compared to the state-of-the-art work where at least one of these key specifications is severely sacrificed in the pursuit of others. A 16×16 sub-array system for in vitro neural front-end recording has been implemented in 0.35 m CMOS and electrical measurement results are presented.
  • Keywords
    neural nets; observability; sensor arrays; CMOS; activity based neural front-end recording system; electrical measurement; neural action potential; neural activity detection; neural recording; observability; sensor arrays; sensor pitch; size 0.35 mum;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el.2011.1966
  • Filename
    6047942