DocumentCode
1352224
Title
Comparison of Neutron Detection Efficiency of a He-3 Counter and a Boron-10 Loaded Liquid Scintillator
Author
Swiderski, Lukasz ; Moszynski, Marek ; Wolski, Dariusz ; Iwanowska, Joanna ; Szczesniak, Tomasz ; Pausch, Guntram ; Plettner, Cristina ; Stein, Juergen ; Schotanus, Paul ; Hurlbut, Chuck ; Szabelski, Jacek
Author_Institution
Soltan Inst. for Nucl. Studies, Otwock-Swierk, Poland
Volume
57
Issue
5
fYear
2010
Firstpage
2857
Lastpage
2861
Abstract
Performance of a He-3 counter and a B-10 loaded liquid scintillator EJ309B5 has been studied in terms of neutron detection efficiency. The measurements were carried out in a mixed field of neutron and gamma radiation from an intense ( ~ 106 neutrons/s/4π)252Cf source. The response of both detectors to background and high intensity gamma radiation ( ~ 100 μSv/h at a detector) from a 60Co source has been measured to establish background count rate and gamma rays cut-off point, respectively. The analysis showed that the properties of a He-3 counter are significantly better than that of EJ309B5. However, it has been pointed out how to improve the performance of a liquid scintillator in order to reduce gamma radiation sensitivity.
Keywords
helium-3 counters; liquid scintillation detectors; neutron detection; 60Co source; Boron-10 loaded liquid scintillator; EJ309B5 liquid scintillator; He-3 counter; background count rate; gamma radiation; gamma radiation sensitivity; gamma ray cut-off point; intense 252Cf source; neutron detection efficiency; neutron radiation; Detectors; Electron tubes; Energy measurement; Neutrons; Radiation detectors; Scattering; Sensitivity; B-10 loaded liquid scintillator; EJ309B5; He-3 counter; detection efficiency; neutrons;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2010.2064790
Filename
5603461
Link To Document