DocumentCode :
1356457
Title :
Time-Resolved Interventional Cardiac C-arm Cone-Beam CT: An Application of the PICCS Algorithm
Author :
Chen, Guang-Hong ; Thériault-Lauzier, Pascal ; Tang, Jie ; Nett, Brian ; Leng, Shuai ; Zambelli, Joseph ; Qi, Zhihua ; Bevins, Nicholas ; Raval, Amish ; Reeder, Scott ; Rowley, Howard
Author_Institution :
Dept. of Med. Phys., Radiol., Human Oncology, Univ. of Wisconsin-Madison, Madison, WI, USA
Volume :
31
Issue :
4
fYear :
2012
fDate :
4/1/2012 12:00:00 AM
Firstpage :
907
Lastpage :
923
Abstract :
Time-resolved cardiac imaging is particularly interesting in the interventional setting since it would provide both image guidance for accurate procedural planning and cardiac functional evaluations directly in the operating room. Imaging the heart in vivo using a slowly rotating C-arm system is extremely challenging due to the limitations of the data acquisition system and the high temporal resolution required to avoid motion artifacts. In this paper, a data acquisition scheme and an image reconstruction method are proposed to achieve time-resolved cardiac cone-beam computed tomography imaging with isotropic spatial resolution and high temporal resolution using a slowly rotating C-arm system. The data are acquired within 14 s using a single gantry rotation with a short scan angular range. The enabling image reconstruction method is the prior image constrained compressed sensing (PICCS) algorithm. The prior image is reconstructed from data acquired over all cardiac phases. Each cardiac phase is then reconstructed from the retrospectively gated cardiac data using the PICCS algorithm. To validate the method, several studies were performed. Both numerical simulations using a hybrid motion phantom with static background anatomy as well as physical phantom studies have been used to demonstrate that the proposed method enables accurate reconstruction of image objects with a high isotropic spatial resolution. A canine animal model scanned in vivo was used to further validate the method.
Keywords :
cardiology; computerised tomography; data acquisition; image reconstruction; medical image processing; numerical analysis; phantoms; PICCS algorithm; canine animal model; cardiac functional evaluation; data acquisition system; high temporal resolution; hybrid motion phantom; image guidance; image reconstruction method; numerical simulations; prior image constrained compressed sensing algorithm; procedural planning; single gantry rotation; time 14 s; time-resolved interventional cardiac C-arm cone-beam computed tomography; Animals; Computed tomography; Data acquisition; Image reconstruction; Phantoms; Three dimensional displays; Cardiac cone-beam CT; compressed sensing; high temporal resolution; interventional imaging; Algorithms; Animals; Computer Simulation; Cone-Beam Computed Tomography; Coronary Angiography; Dogs; Female; Movement; Phantoms, Imaging; Radiographic Image Enhancement; Radiographic Image Interpretation, Computer-Assisted; Reproducibility of Results; Time Factors;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.2011.2172951
Filename :
6056564
Link To Document :
بازگشت