DocumentCode
1356960
Title
Self-Integrating Rogowski Coil for High-Impulse Current Measurement
Author
Metwally, Ibrahim A.
Author_Institution
Dept. of Electr. & Comput. Eng., Sultan Qaboos Univ., Muscat, Oman
Volume
59
Issue
2
fYear
2010
Firstpage
353
Lastpage
360
Abstract
This paper presents the operating principle, the performance under impulse condition, and the design of a self-integrating Rogowski coil for measuring high-impulse currents. Oscillatory and overdamped unidirectional impulse currents are generated up to 10 kA and measured by different methods, namely, two commercial impulse current transformers, a resistive shunt, and the newly designed self-integrating Rogowski coil. For the resistive method, the voltage drop across the shunt affects the total voltage measured across the test object and causes distortion in the case of simultaneous measurement by an impulse current transformer having a small aperture. The Rogowski coil theory under impulse current conditions is presented. The concept of transmission lines is used, and the line parameters are optimized to achieve a self-integrating mode and avoid spurious effects of reflections and stray capacitances. The output voltage linearity of the designed self-integrating Rogowski coil is checked using different linear and nonlinear loads and coil termination resistances. It is found that at a termination resistance of 1 ??, satisfactory impulse current waveforms are measured by taking the commercial impulse current transformer as a reference signal. Results reveal that the measurement errors for the current peak and front and tail times are ??2%, ??6%, and ?? 12%, respectively. Overdamped impulse currents are generated by different generator capacitances, where comparisons between the measured and calculated current waveform parameters have shown good agreement.
Keywords
coils; current transformers; electric current measurement; pulse transformers; coil termination resistances; high-impulse current measurement; impulse current transformers; linear-nonlinear loads; overdamped unidirectional impulse currents; resistance 1 ohm; resistive shunt; self-integrating Rogowski coil; self-integrating mode; transmission lines; voltage drop; Impulse current generator (ICG); Rogowski coil; impulse current measurement; resistive current shunt; transient analysis; transient voltage surge suppressors (TVSSs);
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2009.2023821
Filename
5223579