DocumentCode :
1359952
Title :
Comparison between the ripple technique and the direct calibration method for characterisation of directional devices
Author :
Rodriguez, M.
Author_Institution :
RF Stand. Lab., Inst. Nacional de Tecnica Aeroespacial, Madrid
Volume :
147
Issue :
4
fYear :
2000
fDate :
7/1/2000 12:00:00 AM
Firstpage :
196
Lastpage :
201
Abstract :
A comparison between two different methods for measurement of directivity and test port match of directional couplers and bridges is presented. One of them, the popular ripple technique, is scalar. The other method, the so-called direct calibration method as derived by Juroshek (1997), with slight differences for the simultaneous measurement of both parameters, is vectorial. The two methods make use of a vector network analyser (VNA) and the calibrations performed are verified against known standards. The uncertainty associated with both techniques is computed
Keywords :
calibration; directional couplers; measurement standards; measurement uncertainty; network analysers; bridges; direct calibration; directional couplers; measurement standards; measurement uncertainty; ripple technique; scalar technique; simultaneous measurement; vector network analyser; vectorial technique;