DocumentCode :
1361084
Title :
Noninvasive thermometry assisted by a dual-function ultrasound transducer for mild hyperthermia
Author :
Lai, Chun-Yen ; Kruse, Dustin E. ; Caskey, Charles F. ; Stephens, Douglas N. ; Sutcliffe, Patrick L. ; Ferrara, Katherine W.
Author_Institution :
Dept. of Biomed. Eng., Univ. of California at Davis, Davis, CA, USA
Volume :
57
Issue :
12
fYear :
2010
fDate :
12/1/2010 12:00:00 AM
Firstpage :
2671
Lastpage :
2684
Abstract :
Mild hyperthermia is increasingly important for the activation of temperature-sensitive drug delivery vehicles. Noninvasive ultrasound thermometry based on a 2-D speckle tracking algorithm was examined in this study. Here, a commercial ultrasound scanner, a customized co-linear array transducer, and a controlling PC system were used to generate mild hyperthermia. Because the co-linear array transducer is capable of both therapy and imaging at widely separated frequencies, RF image frames were acquired during therapeutic insonation and then exported for off-line analysis. For in vivo studies in a mouse model, before temperature estimation, motion correction was applied between a reference RF frame and subsequent RF frames. Both in vitro and in vivo experiments were examined; in the in vitro and in vivo studies, the average temperature error had a standard deviation of 0.7°C and 0.8°C, respectively. The application of motion correction improved the accuracy of temperature estimation, where the error range was -1.9 to 4.5°C without correction compared with -1.1 to 1.0°C following correction. This study demonstrates the feasibility of combining therapy and monitoring using a commercial system. In the future, real-time temperature estimation will be incorporated into this system.
Keywords :
biomedical ultrasonics; hyperthermia; medical image processing; physiological models; thermometers; ultrasonic therapy; ultrasonic transducers; 2D speckle tracking algorithm; RF image frames; colinear array transducer; dual-function ultrasound transducer; mild hyperthermia; motion correction; mouse model; noninvasive ultrasound thermometry; real-time temperature estimation; temperature estimation; temperature-sensitive drug delivery vehicles; therapeutic insonation; ultrasound scanner; Arrays; Biomedical imaging; Medical treatment; Temperature measurement; Transducers; Tumors; Algorithms; Animals; Breast Neoplasms; Female; Hyperthermia, Induced; Image Processing, Computer-Assisted; Mice; Neoplasm Transplantation; Phantoms, Imaging; Sepharose; Signal Processing, Computer-Assisted; Temperature; Thermography; Transducers; Ultrasonography;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2010.1741
Filename :
5610553
Link To Document :
بازگشت