Title :
Novel Point-Like
Radioactive Source With Spherical Positron Absorber
Author :
Hasegawa, Tomoyuki ; Oda, Keiichi ; Yamada, Takahiro ; Matsumoto, Mikio ; Sato, Yasushi ; Murayama, Hideo ; Takei, Hideyuki
Author_Institution :
Sch. of Allied Health Sci., Kitasato Univ., Sagamihara, Japan
Abstract :
Purpose: In the conventional techniques for determining the calibration factors of positron emission tomography (PET) scanners, cylindrical water and resin phantoms with radioisotope 18F or 68Ge/68Ga are used. In these methods, however, the results depend on attenuation and scatter correction. The purpose of this study is to develop a point-like 68Ge/68Ga radioactive source that can be used to determine the calibration factors of PET scanners without the uncertainty of attenuation and scatter correction. Methods: A spherical absorber design was employed to realize a symmetric angular distribution of emitted annihilation photons. A Geant4-based Monte Carlo simulation code was used to compare physics characteristics of point-like sources with various absorber materials. On the basis of this simulation, a point-like 68Ge/68Ga source with a spherical aluminum absorber was manufactured. Its radioactivity was calibrated at an accredited national calibration facility. A calibration factor of a clinical PET scanner was then obtained with a point-like source and compared with that obtained by a standard cross-calibration method. Results: The emission probability of 0.511 MeV annihilation photon pairs per positron decay was typically 0.6-0.8. The fraction of background photon pairs was 6-8% in the energy region of 0.4-0.6 MeV. Considering these two figures, lower density materials such as aluminum and pol(ymethyl methacrylate) (PMMA) were preferable. For the aluminum absorber, a diameter of 8 mm was suitable to prevent positrons from escaping. The calibration factor obtained with the point-like source agreed with that obtained by the standard method within 2-3%. Conclusion: A point-like 68Ge/68Ga radioactive source was successfully designed, manufactured, and used for determining a calibration factor of a PET scanner. It can be considered a practical tool for calibrating and ev- luating the quantitative aspects of PET scanners.
Keywords :
Monte Carlo methods; calibration; gallium; germanium; image scanners; positron emission tomography; probability; radioactive sources; radioisotopes; 68Ga; 68Ge; Geant4-based Monte Carlo simulation; PET scanner; PMMA; aluminum absorber; annihilation photon pairs; annihilation photons; attenuation; conventional techniques; electron volt energy 0.4 MeV to 0.6 MeV; electron volt energy 0.511 MeV; emission probability; point-like sources; positron absorber; positron emission tomography scanner; radioactive source; radioisotope; resin phantoms; scatter correction; standard cross-calibration method; symmetric angular distribution; Aluminum; Calibration; Materials; Phantoms; Photonics; Positron emission tomography; Positrons; Calibration; positron emission tomography; standardization;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2011.2166806