• DocumentCode
    1364827
  • Title

    Radiation patterns of dual concentric conductor microstrip antennas for superficial hyperthermia

  • Author

    Stauffer, Paul R. ; Rossetto, Francesca ; Leencini, M. ; Gentilli, Guido Biffi

  • Author_Institution
    Dept. of Radiat. Oncology, California Univ., San Francisco, CA, USA
  • Volume
    45
  • Issue
    5
  • fYear
    1998
  • fDate
    5/1/1998 12:00:00 AM
  • Firstpage
    605
  • Lastpage
    613
  • Abstract
    The finite difference time domain (FDTD) method has been used to calculate electromagnetic radiation patterns from 915-MHz dual concentric conductor (DCC) microwave antennas that are constructed from thin and flexible printed circuit board (PCB) materials. Radiated field distributions are calculated in homogeneous lossy muscle tissue loads located under variable thickness coupling bolus layers. This effort extends the results of previous investigations to consider more realistic applicator configurations with smaller 2 cm-square apertures and different coupling bolus materials and thicknesses, as well as various spacings of multiple-element arrays. Results are given for practical applicator designs with microstrip feedlines etched on the backside of the PCB antenna array instead of previously tested bulky coaxial-cable feedline connections to each radiating aperture. The results demonstrate that for an optimum coupling bolus thickness of 2.5-5 mm, the thin, flexible, and lightweight DCC antennas produce effective heating to the periphery of each aperture to a depth of approximately 1 cm, and may be combined into arrays for uniform heating of large area superficial tissue regions with the 50% power deposition contour conforming closely to the outer perimeter of the array.
  • Keywords
    finite difference time-domain analysis; hyperthermia; microstrip antenna arrays; microwave antenna arrays; microwave heating; muscle; physiological models; radiation therapy; 1 cm; 2 cm; 2.5 to 5 mm; 50% power deposition contour; 915 MHz; array outer perimeter; bulky coaxial-cable feedline connections; dual concentric conductor microstrip antennas; homogeneous lossy muscle tissue loads; medical instrumentation; optimum coupling bolus thickness; radiated field distributions; radiation patterns; superficial hyperthermia; therapeutic heating; thin flexible printed circuit board materials; tissue heating; uniform heating; variable thickness coupling bolus layers; Antenna radiation patterns; Apertures; Biological materials; Conducting materials; Conductors; Coupling circuits; Finite difference methods; Hyperthermia; Microstrip antenna arrays; Microstrip antennas; Electric Conductivity; Electromagnetic Fields; Equipment Design; Hyperthermia, Induced; Microwaves; Models, Biological; Muscle, Skeletal;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.668751
  • Filename
    668751