DocumentCode :
1365263
Title :
A Regulator for Pressure-Controlled Total-Liquid Ventilation
Author :
Robert, Raymond ; Micheau, Philippe ; Avoine, Olivier ; Beaudry, Benoit ; Beaulieu, Alexandre ; Walti, Hervé
Author_Institution :
Mech. Eng. Dept., Univ. de Sherbrooke, Sherbrooke, QC, Canada
Volume :
57
Issue :
9
fYear :
2010
Firstpage :
2267
Lastpage :
2276
Abstract :
Total-liquid ventilation (TLV) is an innovative experimental method of mechanical-assisted ventilation in which lungs are totally filled and then ventilated with a tidal volume of perfluorochemical liquid by using a dedicated liquid ventilator. Such a novel medical device must resemble other conventional ventilators: it must be able to conduct controlled-pressure ventilation. The objective was to design a robust controller to perform pressure-regulated expiratory flow and to implement it on our latest liquid-ventilator prototype (Inolivent-4). Numerical simulations, in vitro experiments, and in vivo experiments in five healthy term newborn lambs have demonstrated that it was efficient to generate expiratory flows while avoiding collapses. Moreover, the in vivo results have demonstrated that our liquid ventilator can maintain adequate gas exchange, normal acid-base equilibrium, and achieve greater minute ventilation, better oxygenation and CO2 extraction, while nearing flow limits. Hence, it is our suggestion to perform pressure-controlled ventilation during expiration with minute ventilation equal or superior to 140 mL· min-1·kg-1 in order to ensure PaCO2 below 55 mmHg. From a clinician´s point of view, pressure-controlled ventilation greatly simplifies the use of the liquid ventilator, which will certainly facilitate its introduction in intensive care units for clinical applications.
Keywords :
biomedical equipment; carbon compounds; gases; medical control systems; pneumodynamics; pressure control; robust control; ventilation; CO2; CO2 extraction; gas exchange; liquid ventilator; lungs; mechanical-assisted ventilation; medical device; normal acid-base equilibrium; oxygenation; perfluorochemical liquid; pressure-controlled total-liquid ventilation; pressure-regulated expiratory flow; robust controller; Liquid-assisted ventilation; liquid ventilator; newborn lambs; perfluorochemical liquid; pressure control; Animals; Computer Simulation; Equipment Design; Fluorocarbons; Liquid Ventilation; Models, Biological; Pressure; Sheep; Tidal Volume;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2009.2031096
Filename :
5233807
Link To Document :
بازگشت