Title :
Development of a compact wavelength-shifting fibre gamma camera
Author :
Soares, A.J. ; Cullum, I. ; Miller, D.J. ; Royle, G.J. ; Speller, R.D.
Author_Institution :
Dept. of Med. Phys. & Bioeng., Univ. Coll. London, UK
fDate :
6/1/2000 12:00:00 AM
Abstract :
A compact wavelength-shifting fibre (WSF) gamma camera for 140 keV gamma rays has been developed and an evaluation of its performance is presented and discussed. A spatial resolution of 4 mm FWHM was achieved which is comparable to that of modern Anger cameras. Simulation results predict a spatial resolution better than 3 mm if the WSF signals are 3 times higher than those obtained experimentally ways of achieving this goal are discussed
Keywords :
biomedical equipment; cameras; gamma-ray detection; image resolution; radioisotope imaging; 140 keV; 3 mm; 4 mm; compact wavelength-shifting fibre gamma camera; medical diagnostic imaging; medical instrumentation; modern Anger cameras; nuclear medicine; spatial resolution; Cameras; Educational institutions; Energy resolution; Gamma rays; High-resolution imaging; Nuclear medicine; Optical imaging; Physics; Prototypes; Spatial resolution;
Journal_Title :