DocumentCode :
1373801
Title :
Shear Wave Velocity Imaging Using Transient Electrode Perturbation: Phantom and ex vivo Validation
Author :
DeWall, Ryan J. ; Varghese, Tomy ; Madsen, Ernest L.
Author_Institution :
Dept. of Med. Phys., Univ. of Wisconsin-Madison, Madison, WI, USA
Volume :
30
Issue :
3
fYear :
2011
fDate :
3/1/2011 12:00:00 AM
Firstpage :
666
Lastpage :
678
Abstract :
This paper presents a new shear wave velocity imaging technique to monitor radio-frequency and microwave ablation procedures, coined electrode vibration elastography. A piezoelectric actuator attached to an ablation needle is transiently vibrated to generate shear waves that are tracked at high frame rates. The time-to-peak algorithm is used to reconstruct the shear wave velocity and thereby the shear modulus variations. The feasibility of electrode vibration elastography is demonstrated using finite element models and ultrasound simulations, tissue-mimicking phantoms simulating fully (phantom 1) and partially ablated (phantom 2) regions, and an ex vivo bovine liver ablation experiment. In phantom experiments, good boundary delineation was observed. Shear wave velocity estimates were within 7% of mechanical measurements in phantom 1 and within 17% in phantom 2. Good boundary delineation was also demonstrated in the ex vivo experiment. The shear wave velocity estimates inside the ablated region were higher than mechanical testing estimates, but estimates in the untreated tissue were within 20% of mechanical measurements. A comparison of electrode vibration elastography and electrode displacement elastography showed the complementary information that they can provide. Electrode vibration elastography shows promise as an imaging modality that provides ablation boundary delineation and quantitative information during ablation procedures.
Keywords :
biological effects of microwaves; biomechanics; biomedical electrodes; biomedical ultrasonics; elastic waves; liver; phantoms; radiation therapy; vibrations; ablation needle; electrode displacement elastography; electrode vibration elastography; ex vivo bovine liver ablation; ex vivo validation; finite element model; microwave ablation; piezoelectric actuator; radiofrequency ablation; shear modulus variations; shear wave velocity imaging; tissue-mimicking phantoms; transient electrode perturbation; ultrasound simulations; Electrodes; Ellipsoids; Materials; Phantoms; Radio frequency; Ultrasonic imaging; Vibrations; Electrode vibration elastography; radio-frequency (RF) ablation; shear wave tracking; time-to-peak; ultrasound; Animals; Cattle; Elasticity Imaging Techniques; Electrodes; Hepatectomy; Image Enhancement; Image Interpretation, Computer-Assisted; Liver; Phantoms, Imaging; Reproducibility of Results; Sensitivity and Specificity; Shear Strength; Surgery, Computer-Assisted; Vibration;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.2010.2091412
Filename :
5625913
Link To Document :
بازگشت