DocumentCode :
1377485
Title :
Fabrication and Measurement of the Main Electrodes of the NMIA-BIPM Calculable Cross Capacitors
Author :
Small, Greig W. ; Fiander, John R.
Author_Institution :
Nat. Meas. Inst. of Australia, Lindfield, NSW, Australia
Volume :
60
Issue :
7
fYear :
2011
fDate :
7/1/2011 12:00:00 AM
Firstpage :
2489
Lastpage :
2494
Abstract :
Two new calculable cross capacitors are being constructed by the National Measurement Institute of Australia and the Bureau International des Poids et Mesures, with the aim of limiting sources of uncertainty to 1 nF/F, where achievable. In particular, the main electrodes, which are four stainless steel bars 470 mm long and 50 mm in diameter, must be geometrically accurate to 100 nm. An automated system based on capacitance probes has been constructed for the purpose of measuring the geometry of the bars with an uncertainty of around 20 nm. Two sets of electrodes have been finished to the required accuracy, although further improvement may be possible.
Keywords :
bars; capacitance measurement; capacitors; electrodes; measurement standards; shape measurement; stainless steel; NMIA-BIPM calculable cross capacitor; automated system; capacitance probe; electrode measurement; size 470 mm; size 50 mm; stainless steel bars; uncertainty sources; Bars; Bridge circuits; Capacitance; Capacitors; Electrodes; Geometry; Probes; Bridge circuits; capacitors; impedance; measurement standards; shape measurement;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2010.2088453
Filename :
5634113
Link To Document :
بازگشت