    An improved configuration for the reduction of EMG in electrode cuff recordings: a theoretical approach

  • Author

    Rahal, Mouhamed ; Winter, Jeff ; Taylor, John ; Donaldson, Nick

  • Author_Institution
    Dept. of Electron. & Electr. Eng., Univ. Coll. London, UK
  • Volume
    47
  • Issue
    9
  • fYear
    2000
  • Firstpage
    1281
  • Lastpage
    1284
  • Abstract
    A theoretical investigation of different electroneurogram recording techniques using electrode cuffs is presented. A new screened tripole arrangement is proposed with a higher inherent signal to interference ratio than the true tripole, which also allows the nulling of the residual electromyogram signal. The reduction in interference is small because the electrode impedance is large compared to the source resistance.
  • Keywords
    biomedical electrodes; electromyography; medical signal processing; neurophysiology; EMG reduction technique; electrode cuff recordings; electrode impedance; electroneurogram; inherent signal to interference ratio; interference ratio; residual electromyogram signal nulling; source resistance; true tripole; Electric resistance; Electrodes; Electromyography; Feedback; Impedance; Insulation; Interference; Neuromuscular stimulation; Signal to noise ratio; Voltage; Biomedical Engineering; Biometry; Electrodes; Electrodiagnosis; Electromyography; Humans; Peripheral Nerves;