An improved configuration for the reduction of EMG in electrode cuff recordings: a theoretical approach
Author
Rahal, Mouhamed ; Winter, Jeff ; Taylor, John ; Donaldson, Nick
Author_Institution
Dept. of Electron. & Electr. Eng., Univ. Coll. London, UK
Volume
47
Issue
9
fYear
2000
Firstpage
1281
Lastpage
1284
Abstract
A theoretical investigation of different electroneurogram recording techniques using electrode cuffs is presented. A new screened tripole arrangement is proposed with a higher inherent signal to interference ratio than the true tripole, which also allows the nulling of the residual electromyogram signal. The reduction in interference is small because the electrode impedance is large compared to the source resistance.
Keywords
biomedical electrodes; electromyography; medical signal processing; neurophysiology; EMG reduction technique; electrode cuff recordings; electrode impedance; electroneurogram; inherent signal to interference ratio; interference ratio; residual electromyogram signal nulling; source resistance; true tripole; Electric resistance; Electrodes; Electromyography; Feedback; Impedance; Insulation; Interference; Neuromuscular stimulation; Signal to noise ratio; Voltage; Biomedical Engineering; Biometry; Electrodes; Electrodiagnosis; Electromyography; Humans; Peripheral Nerves;