Title :
Simplified approach to assessing intravenous flow characteristics in the therapeutic infusion range
Author :
Philip, James H. ; Philip, Beverly K.
Author_Institution :
Dept. of Anesthesia, Harvard Med. Sch., Boston, MA, USA
Volume :
35
Issue :
12
fYear :
1988
Firstpage :
1093
Lastpage :
1094
Abstract :
It is shown that the pressure-flow (P-F) relationship in intravenous infusion systems can be simply described by the flow parameters R L (linear-resistance) and F E (flow nonlinearity). The relative-pressure error due to nonlinear flow is found to be insignificant in the flow range used for therapeutic intravenous infusion therapy (F\n\n\t\t
Keywords :
biological fluid dynamics; patient treatment; flow nonlinearity; fluid resuscitation; intravenous flow characteristics; linear resistance; pressure-flow relationship; relative-pressure error; therapeutic intravenous infusion therapy; Elasticity; Electrons; Force sensors; Hospitals; Injuries; Instruments; Robots; Shoulder; System testing; Viscosity; Infusions, Intravenous; Models, Theoretical; Rheology;