DocumentCode :
1385335
Title :
Subject-Specific Estimation of Central Aortic Blood Pressure Using an Individualized Transfer Function: A Preliminary Feasibility Study
Author :
Hahn, Jin-Oh ; Reisner, Andrew T. ; Jaffer, Farouc A. ; Asada, H. Harry
Author_Institution :
Dept. of Mech. Eng., Univ. of Alberta, Edmonton, AB, Canada
Volume :
16
Issue :
2
fYear :
2012
fDate :
3/1/2012 12:00:00 AM
Firstpage :
212
Lastpage :
220
Abstract :
This paper presents a new approach to the estimation of unknown central aortic blood pressure waveform from a directly measured peripheral blood pressure waveform, in which a physics-based model is employed to solve for a subject- and state-specific individualized transfer function (ITF). The ITF provides the means to estimate the unknown central aortic blood pressure from the peripheral blood pressure. Initial proof-of-principle for the ITF is demonstrated experimentally through an in vivo protocol. In swine subjects taken through wide range of physiologic conditions, the ITF was on average able to provide central aortic blood pressure waveforms more accurately than a nonindividualized transfer function. Its usefulness was most evident when the subject´s pulse transit time deviated from normative values. In these circumstances, the ITF yielded statistically significant reductions over a nonindividualized transfer function in the following three parameters: 1) 30% reduction in the root-mean-squared error between estimated versus actual central aortic blood pressure waveform (p <; 10-4), 2) >;50% reduction in the error between estimated versus actual systolic and pulse pressures ( p <; 10), and 3) a reduction in the overall breakdown rate (i.e., the frequency of estimation errors >;3 mmHg, p <; 10-4). In conclusion, the ITF may offer an attractive alternative to existing methods that estimates the central aortic blood pressure waveform, and may be particularly useful in nonnormative physiologic conditions.
Keywords :
blood; cardiovascular system; haemodynamics; physiology; transfer functions; ITF; central aortic blood pressure waveform; individualized transfer function; nonnormative physiologic conditions; peripheral blood pressure waveform; physics-based model; preliminary feasibility study; subject-specific estimation; swine; Approximation methods; Blood pressure; Particle measurements; Physiology; Pressure measurement; Reflection; Transmission line measurements; Cardiovascular monitoring; central aortic blood pressure; individualized transfer function (ITF); Animals; Aorta; Blood Pressure; Blood Pressure Determination; Feasibility Studies; Humans; Monitoring, Physiologic; Reproducibility of Results; Signal Processing, Computer-Assisted; Swine;
fLanguage :
English
Journal_Title :
Information Technology in Biomedicine, IEEE Transactions on
Publisher :
ieee
ISSN :
1089-7771
Type :
jour
DOI :
10.1109/TITB.2011.2177668
Filename :
6092490
Link To Document :
بازگشت