DocumentCode
1386703
Title
Co-Registration of Interleaved MEG and ULF MRI Using a 7 Channel Low-
SQUID System
Author
Magnelind, Per E. ; Gomez, John J. ; Matlashov, Andrei N. ; Owens, Tuba ; Sandin, J. Henrik ; Volegov, Petr L. ; Espy, Michelle A.
Author_Institution
Appl. Modern Phys. Group, Los Alamos Nat. Lab., Los Alamos, NM, USA
Volume
21
Issue
3
fYear
2011
fDate
6/1/2011 12:00:00 AM
Firstpage
456
Lastpage
460
Abstract
In this paper we report the first co-registered, interleaved measurements of ultra-low field (ULF) magnetic resonance imaging (MRI) and magnetoencephalography (MEG). Interleaved measurements are interesting for the ultimate aim of combining MEG and functional MRI at ULF. The measurement system consisted of 7 channels with second-order gradiometers coupled to low transition-temperature superconducting quantum interference devices (SQUIDs). The ULF MRI was acquired at a measurement field of 94 μT after a pre-polarization in a 30 mT field. Our results show that the two modalities can be performed with interleaved measurements. However, due to transients from the walls of the magnetically shielded room a waiting time of more than 3 s had to be introduced between the MRI protocol and the auditory stimulus for the MEG.
Keywords
SQUID magnetometers; magnetic resonance imaging; magnetoencephalography; MRI; SQUID; low transition-temperature superconducting quantum interference devices; magnetoencephalography magnetic resonance imaging; measurement system; second-order gradiometers; ultra-low field magnetic resonance imaging; Coils; Magnetic noise; Magnetic resonance imaging; Magnetic shielding; Protocols; SQUIDs; Magnetic resonance imaging; magnetoencephalography; microtesla;
fLanguage
English
Journal_Title
Applied Superconductivity, IEEE Transactions on
Publisher
ieee
ISSN
1051-8223
Type
jour
DOI
10.1109/TASC.2010.2088353
Filename
5643164