Title :
A model of the high count rate performance of NaI(Tl)-based PET detectors
Author :
Wear, J.A. ; Karp, J.S. ; Freifelder, R. ; Mankoff, D.A. ; Muehllehner, G.
Author_Institution :
Dept. of Radiol., Pennsylvania Univ., Philadelphia, PA, USA
Volume :
45
Issue :
3
fYear :
1998
fDate :
6/1/1998 12:00:00 AM
Firstpage :
1231
Lastpage :
1237
Abstract :
A detailed model of the response of large area NaI(TI) detectors used in PET and their triggering and data acquisition electronics has been developed, This allows the authors to examine the limitations of the imaging system´s performance due to degradation in the detector performance from light pileup and deadtime from triggering and event processing. Comparisons of simulation results to measurements from the HEAD PENN-PET scanner have been performed to validate the Monte Carlo model. The model was then used to predict improvements in the high count rate performance of the HEAD PENN-PET scanner using different signal integration times, light response functions, and detectors
Keywords :
Monte Carlo methods; biomedical electronics; biomedical equipment; modelling; positron emission tomography; sodium compounds; solid scintillation detectors; thallium; HEAD PENN-PET scanner; Monte Carlo model; NaI(Tl)-based PET detectors model; NaI:Tl; data acquisition electronics; detector performance degradation; event processing; high count rate performance; imaging system´s performance; large area NaI(TI) detectors; light response functions; medical diagnostic imaging; medical instrumentation; nuclear medicine; signal integration times; triggering electronics; Data acquisition; Degradation; Event detection; Head; Monte Carlo methods; Performance evaluation; Positron emission tomography; Predictive models; System performance; Time factors;
fLanguage :
English