• DocumentCode
    1389141
  • Title

    Microchannel plate detectors for 14C autoradiography

  • Author

    Lees, J.E. ; Fraser, G.W. ; Carthew, P.

  • Author_Institution
    Dept. of Phys. & Astron., Leicester Univ., UK
  • Volume
    45
  • Issue
    3
  • fYear
    1998
  • fDate
    6/1/1998 12:00:00 AM
  • Firstpage
    1288
  • Lastpage
    1292
  • Abstract
    We describe a new type of detector for digital autoradiography which combines high sensitivity, useful detecting area (93×93 mm) and good spatial resolution (<100 μm). This instrument is suitable, inter alia, for imaging beta-emitting whole body thin tissue slices. Measured detector sensitivity and linearity derived from images of a 14C Standard are presented. We also describe results from (i) a 14C labelled rat whole body slice showing selective radiopharmaceutical uptake in clearly identifiable organs and (ii) 14C-Tamoxifen doped semi-thin tissue sections of rat womb and liver
  • Keywords
    beta-ray detection; biomedical equipment; microchannel plates; radioisotope imaging; 93 mm; 14C autoradiography; 14C labelled rat whole body slice; 14C-Tamoxifen doped semi-thin tissue section; beta-emitting whole body thin tissue slices; detector sensitivity; digital autoradiography; high sensitivity; linearity; microchannel plate detectors; rat liver; rat womb; selective radiopharmaceutical uptake; spatial resolution; useful detecting area; Anodes; Background noise; Bars; Gamma ray detection; Gamma ray detectors; Glass; Image resolution; Microchannel; Noise level; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.682018
  • Filename
    682018