• DocumentCode
    1390015
  • Title

    Study of bialkali photocathodes below room temperature in the UV/VUV region

  • Author

    Araujo, H.M. ; Chepel, V. Yu ; Lopes, M.I. ; van der Marel, J. ; Marques, R. Ferreira ; Policarpo, A. J P

  • Author_Institution
    Dept. de Fisica, Coimbra Univ., Portugal
  • Volume
    45
  • Issue
    3
  • fYear
    1998
  • fDate
    6/1/1998 12:00:00 AM
  • Firstpage
    542
  • Lastpage
    549
  • Abstract
    Two photomultiplier tubes with bialkali photocathodes and fused silica windows (2" Philips XP2020Q and 1-1/8" Hamamatsu R1668) were studied between room temperature and -160°C, illuminated in dc mode in the UV/VUV wavelength range. The photomultipliers were operated as diodes, with direct measurement of the current of photoelectrons, order to study the effects associated to the photocathode. The influence of light intensity, applied voltage and illuminated area on the response as a function of temperature were investigated. The increase of the electrical resistivity of the emissive layer upon cooling proved to be responsible for a loss of linearity of the photocathode response, leading ultimately to the complete absence of signal at low temperature. A simplified model is presented that accounts for some experimental results. An unexpected increase of response below room temperature was also found at 170 nm
  • Keywords
    electrical resistivity; photocathodes; photomultipliers; 113 to 273 K; Hamamatsu R1668; Philips XP2020Q; UV; VUV; applied voltage; bialkali photocathodes; electrical resistivity; fused silica windows; illuminated area; light intensity; photomultiplier tubes; temperature; Cathodes; Cooling; Current measurement; Diodes; Electric resistance; Photomultipliers; Silicon compounds; Temperature distribution; Voltage; Wavelength measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.682445
  • Filename
    682445