DocumentCode :
1390117
Title :
Studies of ionizing radiation effects on STAR silicon drift detectors
Author :
Bellwied, R. ; Beuttenmuller, R. ; Chen, W. ; DiMassimo, D. ; Dyke, Hazel ; French, A. ; Hall, J.R. ; Hoffmann, G.W. ; Humanic, T.J. ; Kotov, Ivan V ; Kraner, H.W. ; Li, Z. ; Liaw, C.J. ; Lopez, J. ; Lynn, D. ; Rykov, V.L. ; Pandey, S.U. ; Pruneau, C. ; S
Author_Institution :
Wayne State Univ., Detroit, MI, USA
Volume :
45
Issue :
3
fYear :
1998
fDate :
6/1/1998 12:00:00 AM
Firstpage :
623
Lastpage :
625
Abstract :
A 63×63mm rectangular silicon drift detector was irradiated using a 60Co source and its performance was studied. The total accumulated dose was 23.5 krad. The detector performance after room temperature annealing was studied. The detector was found to be sufficiently radiation hard for RHIC applications
Keywords :
annealing; drift chambers; gamma-ray detection; gamma-ray effects; silicon radiation detectors; 23.5 krad; 293 K; 63 mm; RHIC applications; STAR drift detectors; Si; detector performance; gamma ray detection; ionizing radiation effects; rectangular drift detector; room temperature annealing; total accumulated dose; Cathodes; Electron traps; Ionizing radiation; Laboratories; Leakage current; Optical pulse generation; Prototypes; Radiation detectors; Silicon radiation detectors; Temperature;