• DocumentCode
    1390459
  • Title

    Performance comparison of a new approach to nuclear spectral analysis with commercial codes

  • Author

    Reiman, R.T. ; Lasche, G.P. ; Coldwell, R.L. ; Nobel, J.A.

  • Author_Institution
    7284 Cardinal Lane, Niwot, CO, USA
  • Volume
    47
  • Issue
    4
  • fYear
    2000
  • fDate
    8/1/2000 12:00:00 AM
  • Firstpage
    1522
  • Lastpage
    1527
  • Abstract
    The performance of a new approach to nuclear spectral analysis was compared with two widely accepted commercial codes. The comparison measured the new approach´s ability to determine the presence and activity of multiple nuclides in spectra dominated by natural background. The new approach is implemented in the spectral analysis software, RobWinTM. Ten calibration sources ranging in activity from 25 to 450 kBq were placed 20.5 meters from a bare, 100% relative-efficient high purity Ge gamma-ray detector, thus assuring dominance of the natural background radiation over the test nuclides. Accumulated spectra were recorded at ever-doubling time intervals from 225 seconds to 128 hours. All spectra were analyzed by each code according to a defined procedure designed to achieve the best results from the comparator codes. The results from all codes were in general agreement for most nuclides where acquisition times were long, but RobWin´s performance was significantly superior for short counting times
  • Keywords
    calibration; gamma-ray detection; physics computing; radioactive sources; 25 to 450 kBq; Ge; Ge gamma-ray detector; RobWin; calibration sources; commercial codes; nuclear spectral analysis; radioactive source; Calibration; Gamma ray detectors; Germanium; Isotopes; Nuclear measurements; Performance analysis; Radiation detectors; Robustness; Spectral analysis; Testing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.873009
  • Filename