  • Abstract
    Limited count rate ability and open geometry make the image quality obtained from dual head cameras very sensitive to radioactivity outside the camera field of view. The spatial distribution of non true events, scatter and randoms, originating from radioactivity outside the FOV was investigated. The image quality quantified with the noise equivalent count rate, improved by up to 58% as a consequence of shielding. Radioactivity in the first 10 cm outside the FOV was found to contribute most significantly to the detected events. A modular and easily movable patient body shield was designed
  • Keywords
    biomedical equipment; cameras; gamma-ray scattering; positron emission tomography; radioactivity; shielding; 10 cm; dual head coincidence PET camera; image quality; medical diagnostic imaging; medical instrumentation; modular easily movable patient body shield; noise equivalent count rate; nontrue events; nuclear medicine; radioactivity shielding effect; spatial distribution; Cameras; Degradation; Detectors; Event detection; Geometry; Image quality; Magnetic heads; Physics; Positron emission tomography; Scattering;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.873016
  • Filename
    873016