DocumentCode :
1391268
Title :
Limitations on MOSFET dosimeter resolution imposed by 1/f noise
Author :
Tarr, N.G. ; Plett, C. ; Yeaton, A. ; Mackay, G.F. ; Thomson, I.
Author_Institution :
Dept. of Electron., Carleton Univ., Ottawa, Ont., Canada
Volume :
43
Issue :
5
fYear :
1996
fDate :
10/1/1996 12:00:00 AM
Firstpage :
2492
Lastpage :
2495
Abstract :
A sensitive microcomputer-controlled MOSFET-based dosimeter system has been constructed to examine the effects limiting the ability of MOSFET dosimeters to detect very small radiation doses. It is shown that the dose resolution of this system is limited by 1/f noise in the sensor MOSFET´s themselves. Doses below 1 mGy (100 mrad) can be detected. Structural changes to the sensor MOSFET´s that might provide further improvements in resolution through reduction of 1/f noise are discussed
Keywords :
1/f noise; MOSFET; dosimeters; 1/f noise; MOSFET dosimeter resolution; dose resolution; microcomputer-controlled MOSFET-based dosimeter system; very small radiation doses; Electron traps; Ionizing radiation; Ionizing radiation sensors; MOSFET circuits; Noise reduction; Radiation detectors; Sensor systems; Space vehicles; Switches; Switching circuits;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.539397
Filename :
539397
Link To Document :
بازگشت