DocumentCode :
139130
Title :
Fabrication and electrochemical comparison of SIROF-AIROF-EIROF microelectrodes for neural interfaces
Author :
Xiao-Yang Kang ; Jing-Quan Liu ; Hong-Chang Tian ; Bin Yang ; Yanna NuLi ; Chun-Sheng Yang
Author_Institution :
Dept. of Micro/Nano Electron., Shanghai Jiao Tong Univ., Shanghai, China
fYear :
2014
fDate :
26-30 Aug. 2014
Firstpage :
478
Lastpage :
481
Abstract :
Iridium oxide has been widely used in neural recording and stimulation due to its good stability and large charge storage capacity (CSC). In general, the iridium oxide film used in the electrophysiological application can be grouped into three principal classifications: sputtering iridium oxide film (SIROF), activated iridium oxide film (AIROF) and electrodeposited iridium oxide film (EIROF). Although these kinds of iridium oxide all can remarkably reduce the impedance and increase the CSC of the microelectrode, they also exhibit markedly differences in electrochemical performances. After activation, the CSC of EIROF is 68.20 mC/cm2, which is 88.7 % larger than that of the SIROF and 67.6 % larger than that of the AIROF. The impedance at 1 kHz of the three kinds of iridium oxide microelectrode is around 4000 ohm, it is acceptable for the neural interface application. The phase at 1 kHz of the AIROF microelectrode is the largest which is -6.1 degree, about 22.6 % of the SIROF and 44.5 % of the EIROF.
Keywords :
biomedical electrodes; electrochemical electrodes; electrodeposits; microelectrodes; neurophysiology; sputter deposition; thin films; AIROF microelectrode fabrication; EIROF microelectrode fabrication; IrO2; SIROF microelectrode fabrication; activated iridium oxide film; charge storage capacity; electrochemical comparison; electrochemical performances; electrodeposited iridium oxide film; electrophysiological application; iridium oxide microelectrode; neural interfaces; neural recording; neural stimulation; sputtering iridium oxide film; Fabrication; Films; Impedance; Microelectrodes; Platinum; Resists; Sputtering;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2014.6943632
Filename :
6943632
Link To Document :
بازگشت