DocumentCode
1391703
Title
Absolute calibration of a standard temperature noise source for use with s-band radiometers
Author
Hughes, V.A.
Volume
103
Issue
11
fYear
1956
fDate
9/1/1956 12:00:00 AM
Firstpage
669
Lastpage
672
Abstract
To provide a standard temperature noise source for use with S-band radiometers, the noise power from the argon discharge tube CV1881 has been calibrated at 2860 Mc/s using radiometer techniques. The absolute value of the effective temperature of the tube with a discharge current of 180 mA, when mounted across a waveguide parallel to the E-plane and properly matched, is 11 140° K (15.73 dB) with a maximum error of 260° K (0.10 dB), and represents a considerable improvement in accuracy over previous measurements. Since this tube shows a high degree of stability and consistency, it is suggested that it could be used as an absolute standard noise source for the measurement of the noise factor of receivers.
Keywords
microwave measurement;
fLanguage
English
Journal_Title
Proceedings of the IEE - Part B: Radio and Electronic Engineering
Publisher
iet
Type
jour
DOI
10.1049/pi-b-1.1956.0230
Filename
5243007