Title :
Impact of Microwave Pulses on Thermoacoustic Imaging Applications
Author :
Wang, Xiongfei ; Bauer, D.R. ; Vollin, J.L. ; Manzi, D.G. ; Witte, Russell S. ; Xin, Huanhai
Author_Institution :
Dept. of Electr. & Comput. Eng., Univ. of Arizona, Tucson, AZ, USA
fDate :
7/4/1905 12:00:00 AM
Abstract :
Thermoacoustic imaging is a promising modality for breast cancer detection. Acoustic signals generated by tumor targets with different sizes subject to microwave pulses with various widths and waveforms are analytically studied in this letter. Time- and frequency-domain profiles of the acoustic signals are achieved. Some key parameters of the acoustic signals, including peak value and peak-to-peak time interval of the time-domain acoustic pressure and frequency of the global peak and magnitude ratio of the second highest peak to the global peak of the frequency-domain pressure, are discussed. Impact of pulse width on acoustic signal peak-to-peak interval and image spatial resolution is developed. Square pulse is demonstrated to be superior in acquiring higher image resolution and signal-to-noise ratio. This study is helpful to optimizing the microwave pulse, selecting acoustic transducer, and evaluating image resolution.
Keywords :
acoustic imaging; biomedical ultrasonics; cancer; frequency-domain analysis; image resolution; medical image processing; time-domain analysis; tumours; acoustic signals; acoustic transducer; breast cancer detection; frequency-domain pressure; frequency-domain profiles; image spatial resolution; microwave pulses; pulse width; signal-to-noise ratio; thermoacoustic imaging applications; time-domain acoustic pressure; time-domain profiles; tumor targets; Acoustics; Breast cancer; Image resolution; Microwave imaging; Time domain analysis; Tumors; Breast cancer detection; image resolution; pulse waveform; pulse width; thermoacoustic imaging (TAI);
Journal_Title :
Antennas and Wireless Propagation Letters, IEEE
DOI :
10.1109/LAWP.2013.2237743