• DocumentCode
    1393789
  • Title

    A bridge network for the precise measurement of direct capacitance

  • Author

    Lynch, A.C.

  • Volume
    104
  • Issue
    16
  • fYear
    1957
  • fDate
    7/1/1957 12:00:00 AM
  • Firstpage
    363
  • Lastpage
    366
  • Abstract
    For measurements by substitution only, the balance of a transformer bridge network can be made independent of capacitance and conductance from either terminal of the specimen to earth. In the network proposed in the paper both oscillator and detector have one terminal connected to earth. The network is particularly useful for measuring capacitances of less than 100¿¿F in the presence of stray capacitances to earth. It can be used at frequencies between about 4c/s and 20Mc/s; at frequencies up to about 1Mc/s it is suitable for work of the highest precision.
  • Keywords
    bridge instruments; capacitance measurement;
  • fLanguage
    English
  • Journal_Title
    Proceedings of the IEE - Part B: Radio and Electronic Engineering
  • Publisher
    iet
  • Type

    jour

  • DOI
    10.1049/pi-b-1.1957.0167
  • Filename
    5243386