DocumentCode
1394657
Title
Comparisons of image quality factors for phase aberration correction with diffuse and point targets: theory and experiments
Author
Zhao, D. ; Trahey, G.E.
Author_Institution
Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
Volume
38
Issue
2
fYear
1991
fDate
3/1/1991 12:00:00 AM
Firstpage
125
Lastpage
132
Abstract
The use of several alternative image-brightness-based quality factors for phase aberration correction with diffuse and point targets in coherent ultrasonic imaging systems is explored. The factors are similar to the sharpness functions proposed by R.A. Muller et al. (1974) for incoherent imaging systems. Different region of interest (ROI) sizes are used to compare the quality factors. Good agreement is found between theoretical analysis, computer simulations, and phantom experiments. For a point target, the mean of the magnitude of echo signals is inferior to the other studied quality factors in correcting phase aberration. All of the quality factors studied are similar in their ability to correct phase errors for diffuse target images.<>
Keywords
Q-factor; acoustic signal processing; biomedical ultrasonics; computerised picture processing; error correction; speckle; ROI sizes; coherent ultrasonic imaging systems; computer simulations; diffuse targets; image quality factors; medical imaging; phantom experiments; phase aberration correction; phase error correction; point targets; speckle; Brightness; Degradation; Error correction; Frequency; Humans; Image quality; Phased arrays; Q factor; Ultrasonic imaging; Ultrasonic transducer arrays;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.68469
Filename
68469