DocumentCode :
139477
Title :
A low pull-in SU-8 based Capacitive Micromachined Ultrasonic Transducer for medical imaging applications
Author :
Joseph, Jayaraj ; Singh, S.G. ; Vanjari, Siva Rama Krishna
Author_Institution :
Dept. of Electr. Eng., Indian Inst. of Technol. Hyderabad, Hyderabad, India
fYear :
2014
fDate :
26-30 Aug. 2014
Firstpage :
1398
Lastpage :
1401
Abstract :
In this paper we present a thorough analysis of a low pull-in voltage Capacitive Micromachined Ultrasonic Transducer (CMUT) using SU-8 as the membrane material. It is designed to operate at 1 MHz frequency that has a wide range of applications including the imaging of deeper organs. We also propose a simple state-of-the-art fabrication methodology for the same. As compared to the standard Silicon Nitride CMUTs, the proposed structure gives the same electromechanical coupling coefficient with lower membrane dimensions and low pull-in voltage which in turn results in reduced area and power consumption. Moreover the proposed fabrication methodology is a low temperature process which makes it CMOS compatible.
Keywords :
biomedical ultrasonics; ultrasonic imaging; ultrasonic transducers; CMUT imaging; SU-8 based capacitive micromachined ultrasonic transducer; electromechanical coupling coefficient; frequency 1 MHz; medical imaging; membrane material; power consumption; silicon nitride CMUTs; Analytical models; Capacitance; Couplings; Fabrication; Integrated circuit modeling; Materials; Ultrasonic transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2014.6943861
Filename :
6943861
Link To Document :