DocumentCode
1396546
Title
A simple design methodology for elliptical cross-section, transverse, asymmetric, head gradient coils for MRI
Author
Crozier, Stuart ; Doddrell, David M.
Author_Institution
Centre for Magnetic Resonance, Queensland Univ., Brisbane, Qld., Australia
Volume
45
Issue
7
fYear
1998
fDate
7/1/1998 12:00:00 AM
Firstpage
945
Lastpage
948
Abstract
A simple design process for the design of elliptical cross-section, transverse gradient coils for use in magnetic resonance imaging (MRI) is presented. This process is based on a flexible stochastic optimization method and results in designs of high linearity and efficiency with low switching times. A design study of a shielded, transverse asymmetric elliptical coil set for use in neural imaging is presented and includes the minimization of the torques experienced by the gradient set.
Keywords
biomedical NMR; biomedical equipment; coils; efficiency; elliptical cross-section transverse asymmetric head gradient coils; flexible stochastic optimization method; linearity; magnetic resonance imaging; shielded transverse asymmetric elliptical coil set; simple design methodology; simple design process; switching times; torques; Australia; Coils; Design methodology; Inductance; Magnetic heads; Magnetic resonance; Magnetic resonance imaging; Process design; Radio frequency; Torque; Computer Simulation; Equipment Design; Humans; Magnetic Resonance Imaging; Models, Theoretical;