Title :
A very precise short-period comparator for 100 kc/s frequency standards
Author :
Thompson, A.M. ; Archer, R.W.
Author_Institution :
Commonwealth Scientific and Industrial Research Organisation (CSIRO), National Standards Laboratory, Division of Electrotechnology, Sydney, Australia
fDate :
1/1/1959 12:00:00 AM
Abstract :
An instrument is described for the very precise comparison of 100 kc/s frequency standards. A sensitive phase-comparator is combined with a trigger circuit to produce a timing pulse at the instant the two 100 kc/s signals are in phase opposition. The time interval between successive pulses is determined by a counter-type chronometer. For a 10 sec beat period a timing accuracy of 1 microsec is obtained. This corresponds to an accuracy of frequency comparison of 1 in 1013.
Keywords :
frequency measurement; instruments;
Journal_Title :
Proceedings of the IEE - Part B: Radio and Electronic Engineering
DOI :
10.1049/pi-b-1.1959.0015