DocumentCode :
139972
Title :
Reliability of spectral analysis of fetal heart rate variability
Author :
Warmerdam, G.J.J. ; Vullings, R. ; Bergmans, J.W.M. ; Oei, S.G.
Author_Institution :
Fac. of Electr. Eng., Eindhoven Univ. of Technol., Eindhoven, Netherlands
fYear :
2014
fDate :
26-30 Aug. 2014
Firstpage :
2817
Lastpage :
2820
Abstract :
Spectral analysis of fetal heart rate variability could provide information on fetal wellbeing. Unfortunately, fetal heart rate recordings are often contaminated by artifacts. Correction of these artifacts affects the outcome of spectral analysis, but it is currently unclear what level of artifact correction facilitates reliable spectral analysis. In this study, a method is presented that estimates the error in spectral powers due to artifact correction, based on the properties of the Continuous Wavelet Transformation. The results show that it is possible to estimate the error in spectral powers. The information about this error makes it possible for clinicians to assess the reliability of spectral analysis of fetal heart rate recordings that are contaminated by artifacts.
Keywords :
electrocardiography; medical signal processing; obstetrics; signal denoising; spectral analysis; wavelet transforms; Continuous Wavelet Transformation; artifact correction; fetal heart rate recordings; fetal heart rate variability; fetal wellbeing; spectral analysis reliability; spectral powers; Continuous wavelet transforms; Fetal heart rate; Hafnium; Heart rate variability; Reliability; Spectral analysis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2014.6944209