  • Abstract
    A method of computing the signal-to-noise ratio (SNR) in high-frequency magnetic resonance (MR) imaging systems is presented. The method uses a numerical solution to Maxwell´s equations which can capture all relevant electrodynamic effects at high B 0-field strengths. Using this method, the intrinsic SNR of both volume and surface coils loaded with the human head is calculated as a function of frequency. It is shown that although the available SNR from any point scales favorably with frequency, phase inhomogeneity over the volume of the head may pose a challenge to achieving improved SNR with traditional imaging techniques at high B 0-field strengths.
  • Keywords
    Maxwell equations; biomedical MRI; noise; Maxwell´s equations; high B/sub 0/-field strengths; high-frequency magnetic resonance imagers; human head; medical diagnostic imaging; phase inhomogeneity; signal-to-noise ratio computation; surface coils; traditional imaging techniques; volume coils; Coils; Electrodynamics; Frequency; Humans; Magnetic analysis; Magnetic heads; Magnetic resonance; Magnetic resonance imaging; Maxwell equations; Signal to noise ratio; Biomedical Engineering; Biometry; Head; Humans; Magnetic Resonance Imaging; Models, Anatomic; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.880105
  • Filename
    880105