DocumentCode :
1402263
Title :
Coaxial probes with finite ground plane
Author :
Banos-Polglase, J.D. ; Rebollar, Jesus M.
Author_Institution :
ETSI Telecomun. de Madrid.
Volume :
24
Issue :
5
fYear :
1988
fDate :
3/3/1988 12:00:00 AM
Firstpage :
291
Lastpage :
292
Abstract :
The radius of the ground plane of coaxial probes commonly used for the in vivo determination of the dielectric properties of biological tissues is not necessarily large enough to assure its effect is negligible. The authors present a model which allows the determination of the effect of this significant source of error on the impedance of the probe
Keywords :
biomedical electronics; coaxial cables; dielectric measurement; measurement errors; microwave measurement; probes; waveguides; biological tissues measurement; dielectric properties measurement; finite ground plane; ground plane of coaxial probes; in vivo determination; microwave measurement; model; source of error;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
Filename :
5666
Link To Document :
بازگشت