DocumentCode :
1404554
Title :
Simple method for correction of reflection values by means of coaxial quarter wavelength air line impedance standards
Author :
Stumper, Ulrich
Volume :
27
Issue :
5
fYear :
1991
Firstpage :
400
Lastpage :
402
Abstract :
If only impedance standards of low quality are available for the calibration of a reflectometer of a network analyser, a simple correction method can be applied in which complex reflection coefficient values are traced to calculable air line impedance standards. The method has been tested in experiments in 14 mm and 7 mm/N coaxial techniques.
Keywords :
calibration; electric impedance; error correction; measurement standards; network analysers; reflectometers; air line impedance standards; coaxial techniques; complex reflection coefficient values; correction method; error box parameters; network analyser; quarter wavelength; reflection values; reflectometer calibration;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el:19910255
Filename :
64288
Link To Document :