Title :
A novel biphasic-current-pulse calibration technique for electrical neural stimulation
Author :
Maohua Ren ; Jinyong Zhang ; Lei Wang ; Zhenyu Wang
Author_Institution :
Inst. of Biomed. & Health Eng., Shenzhen Inst. of Adv. Technol., Shenzhen, China
Abstract :
One of the major challenge in neural prosthetic device design is to ensure charge-balanced stimulation. This paper presents a new calibration technique to minimize the mismatch between anodic and cathodic current amplitudes. The proposed circuit mainly consists of a digital and an analog calibration, where a successive approximation register (SAR) logic and a comparator are used in digital calibration while a source follower is adopted in analog calibration. With a 0. 18 μm high voltage CMOS process, the simulation shows that the maximum current mismatch is 45 nA (<;0.05%).
Keywords :
bioelectric potentials; calibration; patient treatment; anodic current amplitudes; biphasic-current-pulse calibration technique; cathodic current amplitudes; charge-balanced stimulation; current 45 nA; electrical neural stimulation; high voltage CMOS process; neural prosthetic device design; size 0.18 mum; successive approximation register logic; Calibration; Capacitors; Educational institutions; Electrodes; Integrated circuits; MOSFET; Quantization (signal);
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
DOI :
10.1109/EMBC.2014.6944749