DocumentCode :
1413505
Title :
The ultra-sound image camera
Author :
Smyth, C.N. ; Poynton, F.Y. ; Sayers, J.F.
Volume :
110
Issue :
1
fYear :
1963
fDate :
1/1/1963 12:00:00 AM
Firstpage :
16
Lastpage :
28
Abstract :
The ultra-sound image camera enables a visible image to be reproduced of the ultra-sound intensity distribution on the camera face. Coupled with acoustic lenses, and immersed in a suitable medium, the internal structure of substances or objects transparent to ultrasonic waves can be studied with the instrument. In its uses it is complementary to X-ray and to ultrasonic pulse-echo inspection methods. It has applications to the non-destructive testing of materials and engineering components, and also to medical diagnosis. An ultra-sound camera tube and complete viewing equipment have been built. The sensitivity for unity signal/noise ratio is 10¿7 W/cm2 at the camera face, and the resolution is 0.5 mm at 4 Mc/s. It is one of several possible types of ultra-sound camera. The design is discussed, and directions into which further research and development might be taken are also considered.
Keywords :
instruments; measurement by electrical methods; medical science; testing; ultrasonics;
fLanguage :
English
Journal_Title :
Electrical Engineers, Proceedings of the Institution of
Publisher :
iet
ISSN :
0020-3270
Type :
jour
DOI :
10.1049/piee.1963.0003
Filename :
5247965
Link To Document :
بازگشت