    Characterization of the H3D ASIC Readout System and 6.0 cm ^{3} 3-D Position Sensitive CdZnTe Detectors

  • Author

    Zhang, Feng ; Herman, Cedric ; He, Zhong ; De Geronimo, Gianluigi ; Vernon, Emerson ; Fried, Jack

  • Author_Institution
    Dept. of Nucl. Eng. & Radiol. Sci., Univ. of Michigan, Ann Arbor, MI, USA
  • Volume
    59
  • Issue
    1
  • fYear
    2012
  • Firstpage
    236
  • Lastpage
    242
  • Abstract
    Two 20 mm × 20 mm × 15 mm pixelated CZT detectors made by eV-Products were characterized using the new H3D Application Specific Integrated Circuits (ASIC) readout system developed by the Instrumentation Division at Brookhaven National Laboratory. The ASIC is capable of reading out energy and timing signals from 121 anode pixels and the planar cathode electrode of one CZT detector simultaneously. The system has a measured electronic noise of ~2.2 keV FWHM with a dynamic range from 20 keV to 3.0 MeV. The two detectors achieved energy resolution of 0.48% FWHM and 0.60% FWHM, respectively, at 662 keV for single-pixel events from the entire 6.0 cm3 detection volume at room temperature with an un-collimated 137Cs source. The average (μτ)e of both detectors were measured to be >; 10-2 cm 2/V. The detection efficiency of the two detectors was evaluated at several different energies up to 1.3 MeV by comparing with simulated data. It was found that the total counts agree well between the measured data and the simulated data over the studied energy range. However, the measured photopeak counts were 10-15% lower than simulated photopeak counts at high gamma-ray energies. The analysis shows that the loss of photopeak efficiency is likely due to the charge loss from peripheral pixels to the boundary of detectors.
  • Keywords
    application specific integrated circuits; nuclear electronics; position sensitive particle detectors; readout electronics; semiconductor counters; 3D position sensitive CdZnTe detectors; H3D ASIC readout system; anode pixels; application specific integrated circuits; charge loss; detection efficiency; electron volt energy 20 keV to 3.0 MeV; energy signal read out; peripheral pixels; photopeak counts; pixelated CZT detectors; planar cathode electrode; single pixel events; size 15 mm; size 20 mm; timing signal read out; uncollimated 137Cs source; Anodes; Application specific integrated circuits; Cathodes; Detectors; Energy resolution; Noise; Timing; 3-D; CdZnTe; position sensitive;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2011.2175948
  • Filename
    6122474