DocumentCode :
1415126
Title :
Digital Oscilloscope Calibration Using Asynchronously Sampled Signal Estimation
Author :
Lapuh, R. ; Pinter, B. ; Voljc, B. ; Svetik, Z. ; Lindic, M.
Author_Institution :
Metrol. Inst. of the Republic of Slovenia, Celje, Slovenia
Volume :
60
Issue :
7
fYear :
2011
fDate :
7/1/2011 12:00:00 AM
Firstpage :
2570
Lastpage :
2577
Abstract :
Oscilloscopes are typically calibrated using special signals for each parameter. A common method using only sine-wave signals and analysis of sampled data is proposed to calibrate all major parameters of a digital sampling oscilloscope, greatly simplifying the calibration procedure while attaining lower uncertainties. The procedure can be also fully automated as the readout from the oscilloscope screen is no longer needed.
Keywords :
calibration; digital storage oscilloscopes; signal sampling; asynchronously sampled signal estimation; digital oscilloscope calibration; sampled data analysis; sine-wave signal; Accuracy; Calibration; Frequency estimation; Harmonic distortion; Oscilloscopes; Uncertainty; Bandwidth; calibration; digital oscilloscope; frequency response; harmonic distortions; signal estimation;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2010.2096910