Abstract :
The quantitation of absolute renal functional volume is of interest because many pathological processes are known to affect the growth and parenchymal volume of kidneys. Unlike thresholding and edge-detection methods, the count-based method of volume quantitation is not highly sensitive to boundary definition. In this investigation, count-based kidney volume quantitation was compared to that of fixed threshold-based estimation with and without including attenuation and scatter compensation in the reconstruction. The Alderson Organ Scanning Phantom kidneys were filled with a Tc-99m solution and placed in the Organ scanning phantom. Inserts shaped like the pelvis region of the kidneys were used to vary kidney volume. Emission imaging of the Tc-99m, and transmission imaging of a Gd-153 line source were performed by a triple-headed SPECT system. Without compensating for attenuation and scatter, the count-based method significantly under-estimated kidney volume. When attenuation and scatter compensation were included, count-based volume quantitation was more accurate than volume estimation with a fixed threshold adapted to the image acquisition and reconstruction strategies employed
Keywords :
kidney; medical image processing; single photon emission computed tomography; volume measurement; Gd; Gd-153 line source; SPECT imaging; Tc; Tc-99m solution; attenuation compensation; count-based quantitation; emission imaging; functional renal volume; image acquisition strategy; kidney growth; kidney parenchymal volume; medical diagnostic imaging; nuclear medicine; pathological processes; reconstruction strategy; scatter compensation; transmission imaging; triple-headed SPECT system; Attenuation; Diseases; Image reconstruction; Imaging phantoms; Magnetic resonance imaging; Nuclear medicine; Pathological processes; Scattering; Tomography; Volume measurement;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.708341
Filename :
708341
Link To Document :
بازگشت