• DocumentCode
    1418161
  • Title

    Accurate measurement of millimetre-wave phase-shift from 220 to 325 GHz using dual-channel system

  • Author

    Wu, T.Y.

  • Author_Institution
    Nat. Metrol. Centre, A*STAR, Singapore, Singapore
  • Volume
    48
  • Issue
    1
  • fYear
    2012
  • Firstpage
    31
  • Lastpage
    33
  • Abstract
    A 220-325-GHz millimetre-wave phase-shift measurement system is described. The differential phase-shift measurement uncertainty is estimated to be 0.58°-1.32° for a 0-60 dB rotary vane attenuator (RVA) at 325-GHz. The phase-shift of a WR-03 waveguide shim is measured as a line standard to verify the system. Measurement results of a WR-03 RVA are also presented.
  • Keywords
    measurement uncertainty; millimetre wave measurement; phase measurement; waveguide attenuators; RVA; differential phase shift measurement uncertainty; dual channel system; frequency 220 GHz to 325 GHz; millimetre wave phase shift accurate measurement; rotary vane attenuator; waveguide shim;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el.2011.3351
  • Filename
    6126142