DocumentCode
1420013
Title
A Versatile Wireless Portable Monitoring System for Brain–Behavior Approaches
Author
Chang, Da-Wei ; Liang, Sheng-Fu ; Young, Chung-Ping ; Shaw, Fu-Zen ; Su, Alvin W Y ; Liu, You-De ; Wang, Yu-Lin ; Liu, Yi-Che ; Chen, Jing-Jhong ; Chen, Chun-Yu
Author_Institution
Dept. of Comput. Sci. & Inf. Eng., Nat. Cheng Kung Univ., Tainan, Taiwan
Volume
1
Issue
4
fYear
2011
Firstpage
440
Lastpage
450
Abstract
It is critical to set up a precise and feasible monitoring system for a variety of animal and human studies. A multichannel wireless system for monitoring physiological signals of freely moving rats is presented. This system combines electroencephalogram (EEG) and acceleration signals, enabling the study of association between brain and behavior. A combination of EEG and accelerometers eliminates the necessity for complicated video installation as well as time-consuming and tedious analysis of recorded videos. The IEEE 802.15.4 based wireless communication frees the experimental subject from the hassle of wires and reduces wire artifacts during recording. Long-period continuous recording was possible because of the low power feature of the system. Methods for automatic wake-sleep state discrimination and temporal lobe epileptic seizure detection are also proposed to demonstrate the advantages of the system. An accuracy of up to 96.22% for the automatic discrimination of wake-sleep states is an advantage of our system. In addition, the detection of amygdala-kindling temporal lobe seizures reaches 100% with zero false alarms, greatly saving manpower in the identification of temporal lobe epilepsy.
Keywords
Zigbee; acceleration measurement; accelerometers; electroencephalography; medical disorders; medical signal processing; patient monitoring; sleep; EEG; IEEE 802.15.4 based wireless communication; acceleration signals; accelerometers; amygdala-kindling temporal lobe seizures; automatic wake-sleep state discrimination; brain-behavior approach; electroencephalogram; long-period continuous recording; multichannel wireless system; physiological signals; temporal lobe epilepsy; temporal lobe epileptic seizure detection; versatile wireless portable monitoring system; video installation; video recording; Accelerometers; Biomedical monitoring; Data acquisition; Electroencephalography; Epilepsy; IEEE 802.15 Standards; Wireless communication; Accelerometer; electroencephalogram (EEG); epilepsy; vigilance; wireless;
fLanguage
English
Journal_Title
Emerging and Selected Topics in Circuits and Systems, IEEE Journal on
Publisher
ieee
ISSN
2156-3357
Type
jour
DOI
10.1109/JETCAS.2011.2181454
Filename
6129414
Link To Document